Development of a semi-empirical calibration method by using a LaBr3(Ce) scintillation detector for NORM samples analyses


Published: Apr 19, 2021
E. Ntalla
A. Markopoulos
K. Karfopoulos
C. Potiriadis
A. Clouvas
A. Savidou
Abstract
In this study, a semi-empirical calibration method for NORM samples measurement by using a LaBr3(Ce) scintillator was developed based on a combination of experimental gamma spectrometry measurements and MCNP-X simulations. The aim of this work is to provide us with full energy peak efficiency calibration curves in a wide photon energy range which is of particular importance when selected photon energies of 234Th, 214Pb, 214Bi, 228Ac, 208Tl and 226Ra are to be measured with accuracy.
Article Details
  • Section
  • Poster contributions
References
G. F. Knoll, John Wiley and Sons (1999)
F. Quarati et al., Nucl. Instr. Meth. Phys. Res. A 574, 1 (2007)
K. Ciupek et al., Radioanal. Nucl. Chem. 299, 2 (2014)
W. F. Mueller et al., ANIMMA 2013
G. Gilmore, John Wiley and Sons (2008)