Energy calibration of He-3, He-4 Neutron Proportional Counters
Abstract
Two helium filled proportional counters (He-3, He-4) were irradiated with neu- trons in the energy range 0.025 eV - 10.7 MeV, in the Tandem accelerator, NCSR Demokritos, Athens. The full energy peak in He-3 proportional counter as well as the recoil peaks in He-3 and He-4 counters show linear dependence with neutron energy. A study of gamma ray contribution to the neutron spectra of He-4 counter was performed as a function of the pulse shaping time.
Article Details
- How to Cite
-
Manolopoulou, M., Fragopoulou, M., Stoulos, S., Koukorava, C., & Zamani, M. (2020). Energy calibration of He-3, He-4 Neutron Proportional Counters. HNPS Advances in Nuclear Physics, 13, 222–227. https://doi.org/10.12681/hnps.2972