Position and Energy Resolution of a γ-Camera based on a Position Sensitive Photomultiplier Tube


A. Polychronopoulou
D. Thanasas
N. Giokaris
A. Karabarbounis
D. Maintas
C. N. Papanicolas
E. Stiliaris
Abstract

Studies of the spatial- and energy-resolution of a small field, high resolution γ-Camera system currently being developed in our laboratory are presented here. The system is based on a cylindrical Position Sensitive Photo-Multiplier Tube (HAMAMATSU R2486) with 32 crossed-wired anodes, arranged in two orthogonal groups. The anode outputs are connected to a resistive current divider network and after pre-amplification are guided to a local digitizing system. A PCI-Analog to Digital Converter card with a maximum sampling rate of 20 MHz and the rest of the Data Acquisition System is controlled by software running on the LabVIEW environment. Position and induced charge of the incident light pulses can be easily reconstructed through a variety of offline algorithms.

In the first part of this study the intrinsic response of the PMT and especially its energy and position resolution is presented. The experimental procedure utilizes the controlled pulse light output of a LED guided through fiber glass directly to the PMT’s entrance. The accumulated charge distribution and charge spread as a function of the incident number of photons is studied. In the second part, the response of the integrated γ-Camera system after the application of typical scintillation crystals for 99Tc radioactive phantoms is measured and analyzed.

Article Details
  • Section
  • Oral contributions
References
H.O. Anger, Scintillation Camera, Rev. Sci. Instrum. 29 (1958) 27–33
R. Pani et al. Scintillating array gamma camera for clinical use, Nucl. Instrum. & Methods A 392 (1997) 295–298
J. Kim et al. Development of a miniature scintillation camera using an NaI(Tl) scintillation and PSPMT for scintimammography, Phys. Med. Biol. 45 (2000) 3481–3488
G.K. Loudos et al. A 3D high-resolution gamma camera for radiopharmaceutical studies with small animals Applied Radiation and Isotopes 58 (2003) 501–508
R. Pani et al. A novel compact gamma camera based on flat panel PMT, Nucl. Instrum. & Methods A 513 (2003) 36–41
F. Sanchez, J.M. Benlloch et al. Design and test of a portable mini gamma camera Med. Phys. 31 (2004) 1384–1397
H. Kim et al. SemiSPECT: A small-animal single-photon emission computed tomography (SPECT) imager based on eight cadmium zinc telluride (CZT) detector arrays, Med. Phys. 33 (2006) 465–474
V. Spanoudaki et al. Design and development of a position-sensitive γ-Camera for SPECT imaging based on PCI electronics, Nucl. Instrum. & Methods A 527 (2004) 151–156
V. Spanoudaki et al. A γ-Camera System for SPECT Imaging based on PCI Elec- tronics, MSc Thesis, Physiscs Department, National and Kapodistrian University of Athens, 2003.
Hamamatsu Technical Data Sheet, Position Sensitive Photomultiplier Tubes With Crossed Wired Anodes R2486 Series, Hamamatsu Photonics, 1998.
ADLink Data Sheet, PCI 9812/9810 20 MHz Simultaneous 4-CH Analog Inut Cards, ADLink Technology, 2003.