Induced Radioactivity in Medical Accelerators


Published: Dec 1, 2012
Keywords:
Activation linear accelerator radiation therapy
J. Kalef-Ezra
Abstract

Radiation-induced activation of patients and materials in radiotherapy treatment rooms may result in occupational radiation hazards due to in-room induced activity. Induced activation following 18 MV X-ray irradiations was assessed during commissioning and use of a medical linear accelerator at Ioannina University Hospital. Measurements were carried out using hand-held instruments at twenty locations in the accelerator bunker (three portable dose rate meters, a contamination meter were used). Temporal ambient dose rate data fitted with exponential functions and coupled with in situ spectroscopic data were used to characterize the post-irradiation field and develop detailed and practical site-specific work-plans to keep personnel doses secondary to radiotherapy as low as reasonably practical. The application of the suggested work-plans kept occupational exposure at very low levels during the first 2 y of clinical use of such beams.

Article Details
  • Section
  • Poster contributions
References
IAEA: Handbook on photonuclear data for applications. Cross sections and spectra. IAEA-TECDOC-1178, IAEA, Vienna, 2000
Mao XS., Kase KR. Liu JC. Nelson WR., Kleck JH., Johnsen S.: Neutron sources in the Varian 2100C/2300C medical accelerator calculated by the EGS4 code. Health. Phys. 72: 524-529, 1997
Howell RM., Kry SF., Burget E., Hertel NE, Followill DS.: Secondary neutron spectra from modern Varian, Siemens and Elekta linacs with multileaf collimators. Med. Phys. 36: 4027-4037, 2009
Kalef-Ezra JA. Health physics aspects in treatment rooms after 18 MV X-ray irradiations, Radiat. Prot. Dosim.147: 281-286, 2011
Fisher HW., Tabot V., Poppe B.: Comparison of activation products and induced dose rates in different high-energy medical linear accelerators. Health Phys 94: 272-278, 2008
Rawlinson JA., Islam MK., Galbraith DM.: Dose to radiation therapists from activation at high energy accelerators used from conventional and intensity-modulated radiation therapy. Med. Phys. 29: 598-608, 2002
McGinley PH., Wright BA., Meding CJ.: Dose to radiotherapy technologist from air activation. Med. Phys. 11: 855-858, 2084
Kalef-Ezra J., Bozios G., Tsekeris P., Michalis L:. Patient and personal dosimetry in endovascular brachytherapy with 90Sr/90Y. Radiat. Prot. Dosim. 114: 514-523 2005