A new γ-spectroscopy station at the University of Athens
Published:
Mar 8, 2019
Keywords:
γ-ray spectroscopy HPGe detector efficiency calibration MCNP
Abstract
A new γ-spectroscopy station at the University of Athens has been recently deployed. The station is built around a 25% HPGe detector and aims at supporting the environmental radioactivity studies program of the NuSTRAP group at UoA. The detector needs detailed characterization of its efficiency and calibration over a rather wide range of energies. Besides standard calibration point sources (60Co, 137Cs, 152Eu and 226Ra), detector simulations using the Monte Carlo N-particle transport code (MCNP) were performed. Results of the MCNP calculations are shown and compared with real spectra showing satisfactory agreement
Article Details
- How to Cite
-
Lagaki, V., Kouvaris, E., & Mertzimekis, T. J. (2019). A new γ-spectroscopy station at the University of Athens. HNPS Advances in Nuclear Physics, 22, 126–128. https://doi.org/10.12681/hnps.1919
- Issue
- Vol. 22 (2014): HNPS2014
- Section
- Poster contributions
References
F.B Brown, et al., “MCNP Version5”, LA-UR-02-3199 (2002)
C.A. Kalfas, “SpectrW: a spectra analysis program” (2012), private communication